Model 4900 Series Remote Airborne Particle Counters

4900.jpg (13542 bytes)

0.3 or 0.5µm minimum threshold setting
0.1 or 1.0 cfm flow rate
Two size channels
4–20mA output
Delayed alarm

Connecting remote particle counters to your scada system has never been easier than with Met One’s R4900 Series of miniature particle counters. Their small size allows them to be mounted much nearer to critical monitoring areas at many locations around the cleanroom. All you need to provide is a vacuum source and a communication line. The R4900 Series uses the industry standard 4–20mA interface. Loop current is driven by voltages between 15 and 30 VDC. The sensors use an RS485 serial interface, for programming count time, hold time, alarm level, alarm delay, and scaling for the 4–20mA output. The programming is retained through power off and on cycles. A user-defined alarm level can be set for high counts on channel one. When high counts trigger an alarm, the LED on the counter flashes and an alarm signal is set at the connector. The alarm can be set to trip after the level is exceeded the first time or after the second through ninth time counts exceed the level. The alarm automatically resets at the start of the next count cycle.

4800diag.jpg (10548 bytes)

  Connector (DB-15 Female Connector)
1 Ground
2 Alarm, open drain FET
3 Channel 1 current out
4 Channel 2 current out
5 +24 VDC loop power
6 RS485A
7 N/C
8 RS485B
9 Reserved
10 Reserved
11 Reserved
12 Reserved
13 Ground
14 Reserved
15 +6 VDC Power

Specifications

Size Channels  
R4903: Ch 1/Ch 2 0.3/0.5 micron
R4905, R4915: Ch 1/Ch 2 0.5/5.0 micron
Flow Rate  
R4903, R4905 0.1 cfm
R4915 1.0 cfm
Flow Control critical orifice; requires 18" Hg vacuum
Light Source laser diode (30,000 hours mean time between failure)
Coincidence Loss  
R4903, R4905: 5% at 2x106 particles per cubic foot
R4915: 5% at 4x105 particles per cubic foot
Inlet Pressure ambient to 0.1" Hg vacuum
Indicators power and calibration/alarm LEDs
Power 6 Vdc (+/-10%) at < 250 mA
Loop Supply 15–30VDC
Output 2 size channels; 0mA on sensor fail or no power
Output Scaling program from 10–999,999 counts full scale
Alarm Output open collector FET max sink current (on) 150 mA;
Max V.C. voltage (off) 30 VDC
Alarm Delay 1–9 high count levels
Weight 10.7 ounces (0.30 kg.)
Inlet Port 1/8" ID
Outlet Port 1/4" ID
Environment Operating 55 to 84°F (12 to 29°C), 20 to 95% R.H., noncondensing
Storage -10 to 160°F (-23 to 70°C), up to 98% R.H., noncondensing

Options
Switching power supply (supports up to 16 sensors, 6 VDC, 5 A)
Wall plate power adapter for 8–30 VDC input (use one wall plate per sensor) (can also provide loop current with 15–30 VDC)
Power adapter for 8–30 VDC input (use one per sensor without wall plate)

Ordering Guide
When ordering, specify the following:
Model R4903, R4905, or R4915
Second size channel if different than standard
Options as required