Model 4900 Series Remote
Airborne Particle Counters

0.3 or
0.5µm minimum threshold setting
0.1 or 1.0
cfm flow rate
Two size
channels
420mA
output
Delayed
alarm
Connecting remote
particle counters to your scada system has never been
easier than with Met Ones R4900 Series of miniature
particle counters. Their small size allows them to be
mounted much nearer to critical monitoring areas at many
locations around the cleanroom. All you need to provide
is a vacuum source and a communication line. The R4900
Series uses the industry standard 420mA interface.
Loop current is driven by voltages between 15 and 30 VDC.
The sensors use an RS485 serial interface, for
programming count time, hold time, alarm level, alarm
delay, and scaling for the 420mA output. The
programming is retained through power off and on cycles.
A user-defined alarm level can be set for high counts on
channel one. When high counts trigger an alarm, the LED
on the counter flashes and an alarm signal is set at the
connector. The alarm can be set to trip after the level
is exceeded the first time or after the second through
ninth time counts exceed the level. The alarm
automatically resets at the start of the next count
cycle.

| |
Connector
(DB-15 Female Connector) |
| 1 |
Ground |
| 2 |
Alarm,
open drain FET |
| 3 |
Channel
1 current out |
| 4 |
Channel
2 current out |
| 5 |
+24 VDC
loop power |
| 6 |
RS485A |
| 7 |
N/C |
| 8 |
RS485B |
| 9 |
Reserved |
| 10 |
Reserved |
| 11 |
Reserved |
| 12 |
Reserved |
| 13 |
Ground |
| 14 |
Reserved |
| 15 |
+6 VDC
Power |
Specifications
| Size Channels |
|
| R4903: Ch 1/Ch 2 |
0.3/0.5 micron |
| R4905, R4915: Ch 1/Ch 2 |
0.5/5.0 micron |
| Flow Rate |
|
| R4903, R4905 |
0.1 cfm |
| R4915 |
1.0 cfm |
| Flow Control |
critical orifice; requires 18"
Hg vacuum |
| Light Source |
laser diode (30,000 hours mean time
between failure) |
| Coincidence Loss |
|
| R4903, R4905: |
5% at 2x106 particles per
cubic foot |
| R4915: |
5% at 4x105 particles per
cubic foot |
| Inlet Pressure |
ambient to 0.1" Hg vacuum |
| Indicators |
power and calibration/alarm LEDs |
| Power |
6 Vdc (+/-10%) at < 250 mA |
| Loop Supply |
1530VDC |
| Output |
2 size channels; 0mA on sensor fail
or no power |
| Output Scaling |
program from 10999,999 counts
full scale |
| Alarm Output |
open collector FET max sink current
(on) 150 mA;
Max V.C. voltage (off) 30 VDC |
| Alarm Delay |
19 high count levels |
| Weight |
10.7 ounces (0.30 kg.) |
| Inlet Port |
1/8" ID |
| Outlet Port |
1/4" ID |
| Environment Operating |
55 to 84°F (12 to 29°C), 20 to 95%
R.H., noncondensing |
| Storage |
-10 to 160°F (-23 to 70°C), up to
98% R.H., noncondensing |
Options
Switching power supply
(supports up to 16 sensors, 6 VDC, 5 A)
Wall plate power adapter for
830 VDC input (use one wall plate per sensor) (can
also provide loop current with 1530 VDC)
Power adapter for 830
VDC input (use one per sensor without wall plate)
Ordering
Guide
When ordering, specify the
following:
Model R4903, R4905, or R4915
Second size channel if
different than standard
Options as required
|