met.jpg (12951 bytes)

2800.jpg (9360 bytes)

• 0.3 micron at 1.0 cfm (2813)
• 0.5 micron at 1.0 cfm (2815)
• Counts 4 sizes simultaneously
• Stores up to 500 samples
• Fed-Std-209E calculations
• Integrated bar code entry
• Portability–only 14 lbs (6.4 Kg)

The Model 2813 and 2815 counters provide ease of validation for controlled and critical areas in a variety of industries. The unique bar code entry allows the operator to pre-configure the sampling strategy at each sample point. By posting a bar code label near the sample point, a location marker and a consistent sample process will always be executed. Operators with minimal training can carry out proper monitoring.

Password protection for the configuration section assures consistency of data. Another advantage of this instrument is the bright LED display for easy reading in a package weighing only 14 pounds (6.4 kilograms). Yet these counters provide a full 1.0 cfm (28.3 lpm) flowrate for fast and accurate sampling of airborne particles.

The Model 2813 and 2815 store up to 500 records in the buffer, which can then be transferred to an external printer or downloaded into a computer. For areas with high concentrations of airborne particles, the optional Isodiluters™ can provide dilution at a rate of either 10x or 50x.

The lightweight design of the Model 2813 and 2815 is further enhanced by the sealed enclosure (no exhaust fan), permitting easy wipe-down and virtually particle-free operation.

The Model 2813 and 2815 will provide accurate sampling of four size channels at 1.0 cfm in the most demanding of environments. These counters are "heavyweights" in performance, but light where it counts – in your hands.

Specifications
Size Channels (µm) 4
2813 0.3, 0.5, 5.0, 10.0
2815 0.5, 5.0, 10.0, 25.0
Flow Rate 1.0 cfm (28.3 lpm)
Locations Labels 0 to 999, numeric
Data Storage 500 samples
Sample Time 1 second to 100 minutes
Delay Time 1 second to 100 minutes
Count Cycles 0 to 99
Display Red LED, 7 digits
Count Data Total counts, counts/ft3, counts/m3
Password Numeric, 4 digits
Coincidence Loss Less than 5% at 500,000 particles/ft3
Light Source Laser diode
Printer External
Operator Entry Bar code pen, RS232; set up manual or automatic sampling routines
Vacuum Source Internal DC pump, 50% duty cycle
Power 90-132 VAC or 180-250 VAC, 50/60 Hz
Size 12.5" wide x 4.3" high x 11.6" deep (31.8 x 10.9 x 29.5 cm)
Weight 14 lbs (6.4 kg)
Output RS232C for computer or printer
Accessories Included Bar Code Pen; Isokinetic Probe with Tripod; Purge Filter; AC Cord; Operator’s Manual

When ordering, specify:
Basic Sensitivity 0.3 micron (2813) or 0.5 micron (2815)
Power (AC) 90-132VAC or 180-250VAC

Optional Accessories
Relative Humidity/Temperature Sensor combined with Isokinetic Probe
Filter Scanning Probe, non-electronic
High Pressure Diffuser
Printer, external
Isodiluter™, 10x or 50x
Particle Vision™ PortAll™ v.2 Software
Soft Carrying Case

| Met One Product Menu  | Email INAVAJR  | Main Page |