Model 2432 Particle Counting Manifold

2432.jpg (12647 bytes)

Random location sequencing
High flow purge through all samples
Meets applicable Fed-Std-209E requirements

Met One’s manifold system provides an economical approach to facility monitoring by allowing a single particle counter to monitor up to 16 or 32 sampling locations (Model A2431 or A2432). This system may be connected to a variety of different Met One particle counters, your choice depending on the particle sizes you need to monitor. As a stand-alone system, the particle counters can store and print particle count data (by location), alarm conditions, and Fed-Std-209E statistical calculations.

The manifold system connects to a facility monitoring system using the standard communications port on the particle counter. Met One’s Class One/Windows™ software package can control the manifold, acquire and store data, provide graphic displays and trends, and generate reports. Ease of programming saves setup time, and will give you the flexibility to change your monitoring system as your needs change. For example, the programmed cycle may include up to 100 different sequence steps, and each sampling location may have unique alarm levels based on the particle contamination class at that location. Alarms can be audible or visible at the particle counter or both, or a signal may be generated for remote alarms. The manifold system uses a high-volume pump to assure continuous flows of 3 cfm or more through all ports, so that particles are not lost due to fallout or surface adhesion. House vacuum may be used if it is adequate to achieve the necessary flow rates through the sampling tubes.

2432diag.jpg (15747 bytes)

Specifications

Number of Sampling Ports  
A2431 16
A2432 32
Sample Tubing 1/2" O.D.; 3/8" I.D. (Manifold to up to 98' sampling point)
Max Distance to Particle Counter 50'
Communication Protocol RS232
Controller Size 7.4" X 4.4" X 2.8"
Controller Weight 1.5lbs
Alarm Module each alarm module provides contacts for 16 ports and t/rh;
two alarm modules may be connected for 32-port systems
209E Calculations calculations on stored data, stores 400 data points
Program Cycle up to 100 steps
Power 100V, 115V, 220V, 50/60Hz

Ordering Guide

For 16 locations: Model A2431
For 32 locations: Model A2432

Manifold Controller
Manifold Control Cable, 10'
Counter to Controller Cable, 10'

Particle Counter (choose model)
A2408 (0.3 or 0.5 micron)
A2200B (0.2 micron)
A2100B (0.10 micron)

Vacuum Pump
Vacuum Line (pump/blower to manifold); specify length up to 50'
Sample Tubing (3/8" I.D., 1/2" O.D.) specify length